Time-Tech Spectra · Singapore

FP100Nanosecond Laser Flash Photolysis System

Measure longer-lived transient absorption after pulsed excitation. FP100 combines a nanosecond laser, xenon probe source and automated detection for spectral and kinetic studies.

Time-Tech Spectra FP100 instrument

Is FP100 right for your experiment?

For photochemistry and materials experiments where nanosecond-to-millisecond dynamics matter more than femtosecond resolution.

Measurement
Laser flash photolysis
Detection
200-900 nm configuration
Time window
<20 ms, rate-dependent
Extensions
NIR / MIR / gated spectra

Applications & suitable samples

Photochemistry

Investigate long-lived excited states and reaction intermediates with wavelength-resolved kinetic measurements.

Semiconductor & nanoparticle samples

Follow transient optical signals in visible or optional NIR detection bands.

Long-lifetime emission

Discuss the fluorescence-lifetime measurement path and subtraction of fluorescence background from TA data.

Explore perovskite measurement selection

Specifications & configuration

Manufacturer-listed values below distinguish the measurement parameter from its configuration conditions. The final quotation defines the supplied system.

FP100 specification guide
ParameterPublished value / capabilityConfiguration condition
Probe source75 W xenon lampSource spectral range and system detection range are not the same.
Detection band200-900 nmVisible photodiode: 400-900 nm; detector configuration applies.
GeometryTransmissionConfirm optical path and sample-holder requirements.
Sensitivity≤0.2 mODManufacturer specification; measurement conditions apply.
Temporal resolution≤10 nsNot the same parameter as the full system IRF.
Instrument response (IRF)≤50 ns, typical, on the current supplier webpageThe supplied 2026 catalogue lists ≤20 ns typical. Confirm the response of the quoted laser and detection path.
Time window<20 msDepends on laser repetition rate.
Oscilloscope500 MHz bandwidthListed kinetic-acquisition configuration.
Excitation wavelengths1064 / 532 / 355 / 266 nmSelect the supplied laser and harmonic outputs; not all outputs assumed included.
Spectrometer200 mm monochromator; visible PMT and photodiode optionsDetector coupling depends on the acquisition mode.
AutomationMonochromator, oscilloscope and optical-path switchingAcquisition and analysis software includes background subtraction.

The published 185-2000 nm source spectral range does not mean continuous detection across that range. Optional detection bands and excitation tuning must be quoted separately.

Source: Time-Tech Spectra FP100 and the supplied 2026 catalogue. Reviewed 7 October 2026.

Options & measurement paths

Select the options needed for your measurement; listed choices are not automatically included.

NIR and MIR detection

Extend detection beyond the visible configuration for longer-wavelength transient signals.

NIR and MIR detection configuration
NIR module900-1600 nm.
MIR module1-12 μm listed as an expansion; specify required band and complete optical path.
Tunable OPO excitation

Select excitation wavelengths beyond the fixed laser harmonics.

Tunable OPO excitation configuration
Listed tuning bands210-350, 420-700 and 720-2300 nm in the supplier specification table.
ConfirmThe supplier prose gives 420-680 nm for one band. Obtain a confirmed tuning and pulse-energy specification for the ordered laser.
Gated spectral acquisition

An optional intensified camera acquires spectra at selected delays instead of scanning each wavelength sequentially.

Gated spectral acquisition configuration
SelectionChoose around spectral coverage, gating, signal level and the required acquisition speed.
TimingA camera gate specification alone does not establish the complete experiment's time resolution.

Measurement examples

Original Time-Tech Spectra figures, not HX measurements. Axes, scales and plotted data are retained. Example performance depends on the sample and acquisition conditions.

TPP · transient spectra and recovery

The map, spectral slices and kinetic curve show complementary views of the transient response. The sample and excitation below follow the labels within the original figure.

Sample labelled TPP · 532 nm excitation. Concentration, solvent and pump fluence are not stated in this figure.

FP100 TPP transient absorption map, spectra and kinetic trace
FP100 TPP transient absorption map, spectra and kinetic trace

Figure source: Time-Tech Spectra FP100 measurement examples.

Plan your configuration

Information to prepare for FP100
Sample preparationProvide sample form, solvent/substrate, absorption band and optical density. Confirm cell path length and any air-sensitive handling.
Excitation & detectionDefine the pump wavelength, target probe band and longest useful delay. Check fluorescence interference and the need for NIR or gated spectral acquisition.
InstallationRequest the complete laser/instrument layout, utility needs and appropriate laser-safety provisions before site preparation.

Selection questions

How is FP100 different from TA-LITE?

FP100 targets nanosecond and longer-lived transient processes using flash photolysis. TA-LITE is an ultrafast pump-probe platform. Match the instrument response and time window to the dynamics of interest.

Are fluorescence and transient absorption the same measurement?

No. Emission can contribute background during a TA experiment. Specify whether you need TA, fluorescence lifetime or both, and establish the relevant detection and subtraction workflow.

Compare TA-PRO, TA-LITE and FP100

Brochure & Singapore enquiries

Send HX your sample, target measurement and configuration requirements. We will help identify the questions to resolve with Time-Tech Spectra.

FP100 - 2026 catalogue excerpt. Original catalogue pages 17-19 (printed pages 12-14). Source pages are reproduced unchanged.

You can also contact sales@hxtech.sg.